Calculation of the temperature drop for high-temperature fixed points for different furnace conditions
Ver/ Abrir
Registro completo
Mostrar el registro completo DCFecha
2011-08Derechos
© 2011, Springer Science Business Media, LLC. This version of the article has been accepted for publication, after peer review (when applicable) and is subject to Springer Nature's AM terms of use, but is not the Version of Record and does not reflect post-acceptance improvements, or any corrections. The Version of Record is available online at: ttps://doi.org/10.1007/s10765-011-1019-6
Publicado en
International Journal of Thermophysics, 2011, 32(7-8), 1773 - 1785
Editorial
Springer
Enlace a la publicación
Palabras clave
Back wall
Blackbody tube
Crucible
Fixed point
Furnace temperature profile
Resumen/Abstract
High-temperature fixed points (HTFPs) based on eutectic and peritectic reactions of metals and carbon are likely to become, in the near term, reference standards at high temperatures. Typically for radiation thermometry applications, these HTFPs are generally formed of a graphite crucible, with a reentrant well, an included 120° cone, and a nominal aperture of 3 mm. It is important to quantify the temperature drop at the back wall of the cavity, and to understand the influence of the crucible configuration and furnace conditions on this drop. In order to study these influences, three different situations have been modeled by means of the finite volume method for numerical analysis. The first investigates the influence of the furnace temperature profile on the temperature drop by simulating four different furnace conditions. The other two study variations in the crucible configuration, namely, the thickness of the graphite back wall and the length of the blackbody tube.
Colecciones a las que pertenece
- D19 Artículos [100]