Prioritizing patient safety: analysis of the procurement process of infusion pumps in Spain
Ver/ Abrir
Registro completo
Mostrar el registro completo DCAutoría
Herrero Urigüen, Laura; Sánchez Santiago, María Blanca

Fecha
2023-12Derechos
© 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Publicado en
International Journal of Environmental Research and Public Health, 2023, 20(24), 7179
Editorial
MDPI
Palabras clave
Procurement
Patient safety
Usability
Infusion pumps
Resumen/Abstract
To understand whether patient safety and human factors are considered in healthcare technology procurement, we analyzed the case of infusion pumps as their use critically affects patient safety. We reviewed infusion pump procurements in the Spanish Public Sector Procurement Database. Sixty-three batches in 29 tenders for supplying 12.224 volumetric and syringe infusion pumps and consumables for an overall budget of EUR 30.4 M were identified and reviewed. Concepts related to "ease of use" were identified in the selection requirements of 35 (55.6%) batches, as part of the criteria for the selection of pumps in 23 (36.5%) batches, related to "intuitiveness" in the selection requirements of 35 (55.6%) batches, and in the criteria in 10 (15.9%) batches. No method to evaluate the ease of use, intuitiveness, or usability was mentioned. A review of the procurement teams responsible for the evaluation of the tenders showed no reported human factors or patient safety expertise. We conclude that infusion pump procurement considers usability as a relevant criterion for selection. However, no human factor experts nor specific methods for evaluation of the technology in this field are usually defined. Potential room for refining the selection of healthcare technology to improve patient safety is detected.
Colecciones a las que pertenece
- D08 Artículos [95]
- IDIVAL Artículos [864]