Body temperature measurement uncertainty arising from ear canal geometry and temperature gradients
Ver/ Abrir
Registro completo
Mostrar el registro completo DCFecha
2022-12Derechos
Attribution-NonCommercial-NoDerivatives 4.0 International
Publicado en
Measurement: Sensors, 2022, 24, 100417
Editorial
Elsevier
Enlace a la publicación
Palabras clave
Body temperature
Infrared thermometer
Tympanic membrane
Ear canal
Ear thermometer
Resumen/Abstract
We report the results of a numerical thermal model of the ear canal and tympanic membrane. The model was used to assess the uncertainty contributions in determining body temperature arising from lower ear canal temperature gradients. We find that for reasonable assumptions of such gradients the standard uncertainty contribution is < 0.1 °C.
Colecciones a las que pertenece
- D19 Artículos [94]