In situ refraction index of liquid measurement using polymer optical fibers
Ver/ Abrir
Identificadores
URI: http://hdl.handle.net/10902/2346DOI: 10.1117/12.566697
ISSN: 1996-756X
ISSN: 0277-786X
Registro completo
Mostrar el registro completo DCAutoría
Lomer Barboza, Mauro Matías



Fecha
2004-06-09Derechos
© 2004 Society of Photo-Optical Instrumentation Engineers. One print or electronic copy may be made for personal use only. Systematic electronic or print reproduction and distribution, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited.
Publicado en
Proceedings of SPIE, 2004, vol. 5502, 325-328
Second European Workshop on Optical Fibre Sensors, Santander, 2004
Editorial
SPIE Society of Photo-Optical Instrumentation Engineers
Enlace a la publicación
Palabras clave
Liquid refractive index measurement
Polymer optical fiber
Resumen/Abstract
A device for the measure of index of refraction of liquids in situ based on polymer optical fibers (POFs) was demonstrated. It consists in a sensor head of three passive POFs where two are coupled to two detectors and an electronic unit for the differential measure of signals. A differential operating principle is utilized to reduce noise such as light intensity fluctuation. The device was successfully checked measuring refraction index changes of the water with different concentrations of sugar, salt and alcohol.